Abstract
Method: This research is a descriptive quantitative study with an observational approach which aims to analyze CTDIvol and DLP values based on retrospective data on non-contrast brain CT scans at the Radiology Installation at Hospital X, Central Jakarta.
Results: Calculation of the 2nd quartile (50th percentile) value of CTDIvol and DLP in Adult Non-Contrast Brain CT Scan examinations for the January-December 2022 Period at the Radiology Installation at X Hospital, Central Jakarta is CTDIvol 55.51 mGy and DLP 867.00 mGy* cm.
Conclusion: The 50 percentile value of CTDIvol and DLP in the Adult Non-Contrast Brain CT Scan examination for the January-December 2022 period at the Radiology Installation at X Hospital, Central Jakarta is in accordance with the standards set by BAPETEN/I-DRL 2021. Where is the value for CT Non-Contrast Head Scan, namely a CTDIvol value of 60 mGy and a DLP value of 1275 mGy*cm.