Abstract
This system uses detection of two sensors to measure oxygen saturation in the blood, a microcontroller to process data, a monitor to display data, a buzzer as a warning of the lower limit of measurement and a micro sd card module to store measurement data that has been running. Parameters displayed include the date of measurement, patient ID, SpO2-1, SpO2-2, last SpO2-1 and last SpO2-2. This system will measure and display the SpO2 value of both right and left arms simultaneously. Stroke patients who experience muscle weakness in one hand may experience different values with a hand that does not experience muscle weakness. This study aims to develop an SpO2 monitoring system that will be used to detect stroke patients with non-invasive methods.
The results show that the system has been developed and can be used to measure the patient's SpO2 from both hands simultaneously with a measurement error rate of ±2% for each sensor from standard medical equipment.